D.-W. Seo, S.-T.Khang, S.-C. Chae, J.-W. Yu, G.-K. Lee, and W.-S. Lee
Microwave and Optical Technology Letters, Vol. 58, No. 6, pp. 1271-1275
An open-loop self-adaptive impedance-matching system keeps magnetic-resonance power transfer to implanted biosensors stable despite distance and misalignment, with up to 47% efficiency.
An open-loop self-adaptive wireless power transfer system using magnetic resonance coupling for implantable biosensors, which provides stable wireless power regardless of variations in operating distance and lateral misalignment between the transmitting and receiving antennas, is proposed. According to adaptive impedance matching algorithms under the minimum reflected power conditions, the proposed system achieves an improved power transfer efficiency of maximum 47% within the operating distance.
반사 전력이 최소가 되도록 임피던스를 스스로 맞추는 개루프 무선전력전송 시스템으로, 체내 이식 센서까지의 거리·어긋남이 바뀌어도 최대 47% 효율로 안정적으로 전력을 보낸다.
An open-loop self-adaptive wireless power transfer system using magnetic resonance coupling for implantable biosensors, which provides stable wireless power regardless of variations in operating distance and lateral misalignment between the transmitting and receiving antennas, is proposed. According to adaptive impedance matching algorithms under the minimum reflected power conditions, the proposed system achieves an improved power transfer efficiency of maximum 47% within the operating distance.
The abstract is from the published paper (doi:10.1002/mop.29796).초록은 출판된 논문(doi:10.1002/mop.29796)에서 옮겼습니다.